NIST Temperature Scale Database (SRD 60), Version 3.0

ITS-90 Primary Fixed Points:


Temperature Scale Points
Primary Fixed Points

International Temperature Scale Primary Fixed Points

We provide the primary (defined) fixed points on the International Temperature Scale of 1990 (ITS-90) along with uncertainties. We provide the primary references for all primary fixed points. In addition, to this primary reference, we also provide OTHER references (generally multiple) to the literature for these primary fixed points in the menu item References for Temperature Points.

Instruction:Total Records:   17
Substance
 
Transition
 
T90 (K)
 
u(T90)
(mK)
 
T-T90
(mK)
 
u(T-T90)
(mK)
 
Squib
 
More
Data
 
HeVP0.65 to 5.0<1|T-T90|<2<1manage_search
H2TP13.80330.150.360.01manage_search
e-H2VP117.0350.150.290.01manage_search
e-H2VP220.270.150.160.01manage_search
NeTP24.55610.09-0.060.2manage_search
O2TP54.35840.06-2.210.14manage_search
ArTP83.80580.06-4.210.15manage_search
HgTP234.31560.2-2.890.13manage_search
H2OTP273.1600.10.1manage_search
GaMP302.91460.23.840.34manage_search
InFP429.74850.810.10.8manage_search
SnFP505.0780.611.51.3manage_search
ZnFP692.6770.813.86.9manage_search
AlFP933.473228.76.6manage_search
AgFP1234.93446.214manage_search
AuFP1337.332539.920manage_search
CuFP1357.772552.120manage_search